Daily operation of laminar flow hoods
2025-06-20
Daily operation of laminar flow hoods
Laminar flow hoods are air purification units that provide a local clean environment and are often used in industries with high cleanliness requirements such as medicine, food, and electronics. The following are the daily operation norms and precautions for laminar flow hoods to ensure their normal operation and cleanliness effect:I. Preparations before operation
Environmental inspection
Make sure there is no obvious accumulation of dust or debris around the laminar flow hood to avoid affecting the airflow organization.
Check whether the doors and Windows of the area where the laminar flow hood is located are closed to reduce the entry of polluted air from the outside.
Equipment appearance inspection
Check whether the laminar flow hood shell and control panel are intact, without any damaged or loose parts.
Check whether there is dust accumulation or damage at the air outlet of the high-efficiency filter. If any abnormality is found, it needs to be replaced or cleaned in time.
Inspection of power supply and control system
Confirm that the power connection is normal and that the indicator lights on the control panel show no faults (such as power indicator lights, operation indicator lights, etc.).
Test whether the functions of the control panel buttons (such as start and stop, wind speed adjustment, etc.) are sensitive.
Ii. Startup Operation Process
Connect the power supply.
Turn on the main power switch of the laminar flow hood and wait for the control system to initialize (about 10 to 30 seconds).
Start the fan
Press the "Start" button, and the fan will begin to operate. Observe whether there is a stable airflow at the air outlet (which can be tested through a ribbon or an anemometer).
Adjust the wind speed.
According to the process requirements, the wind speed is adjusted through the control panel (the wind speed range of the laminar flow hood is usually 0.3-0.6m/s) to ensure uniform and stable airflow.
Operation preheating
It is recommended to let the laminar flow hood idle for 15 to 30 minutes after startup to achieve a stable and clean state inside. If necessary, ultraviolet lamp disinfection can be used in combination (if the equipment is equipped).
Iii. Operation and Monitoring during Operation
Real-time monitoring indicators
Wind speed: Use an anemometer to regularly check the wind speed at the air outlet (it is recommended to do so every two hours). If the wind speed is lower than 80% of the set value, it may be that the filter is clogged. Record it promptly and arrange for replacement.
Cleanliness: The number of suspended particles in the laminar flow hood is detected by a dust particle counter to ensure compliance with cleanliness level requirements (such as ISO 5 or higher).
Noise and vibration: The noise during operation should be controlled below 65-75dB. If abnormal vibration or noise occurs, stop the machine immediately to check whether the fan or components are loose.
Material and personnel operation norms
Material transfer: Items must enter through the dedicated transfer window in the clean area to avoid direct introduction of contamination from non-clean areas.
Personnel movements: Operators should wear clean suits, gloves, and other protective equipment. When moving inside the laminar flow hood, it is recommended to do so slowly and minimize frequent limb movements to avoid disrupting the airflow.
Temporary shutdown handling
If the fan is temporarily suspended during use (such as for a short break), it can be kept running at a low speed to avoid frequent starts and stops that may affect the service life of the equipment. Before reusing it, it needs to idle for another 10 to 15 minutes.
Iv. Shutdown Operation Process
Shut down the function module.
First, turn off the ultraviolet lamp (if in use), then press the "Stop" button to turn off the fan.
Cut off the power supply.
Turn off the main power switch and confirm that all the indicator lights on the control panel are off.
Cleaning and Record-keeping
Use a lint-free cloth dipped in 75% ethanol to wipe the surface of the laminar flow hood, the workbench surface, and the surrounding areas to remove residual stains.
Fill in the "Laminar Flow Hood Operation Record", recording the startup time, shutdown time, operation status, wind speed data, abnormal situations, etc.
V. Daily Maintenance and Precautions
Cleaning frequency
Daily: Wipe off surface dust and contaminants, and check whether the differential pressure gauge of the filter (if any) is within the normal range.
Weekly: Clean the primary filter (washable type). If it is a disposable filter, replace it directly.
Monthly: Test the resistance of the high-efficiency filter. When the resistance exceeds 1.5 to 2 times the initial value, replacement should be arranged.
Safety Warning
High-temperature equipment (such as alcohol lamps) must not be placed in the laminar flow hood to avoid damaging the filter or causing a fire.
Non-professionals are not allowed to disassemble the control panel or replace the filter without authorization. Maintenance must be carried out by professional technicians.
If any malfunction occurs during the operation of the equipment (such as abnormal noise from the fan or sudden drop in wind speed), the machine should be stopped immediately and a "Fault" sign should be hung. It can only be used after the problem has been repaired.
Vi. Special Scenario Operations (Such as Disinfection)
Ultraviolet lamp disinfection
When the machine is shut down, turn on the ultraviolet lamp and let it shine for no less than 30 minutes. During disinfection, personnel must leave the site to avoid skin or eye burns caused by ultraviolet rays.
Chemical disinfection
Hydrogen peroxide or formaldehyde can be used for fumigation (by the disinfection procedures for clean areas). After disinfection, adequate ventilation is required. Only when the residual concentration of the gas meets the safety standards can it be used?
Summary
The core of the daily operation of laminar flow hoods lies in "standardized start-up - real-time monitoring - timely maintenance". Through standardized procedures, it can ensure that it continuously provides a stable clean environment. During the operation process, data must be strictly recorded. Any abnormalities found should be dealt with promptly to meet the cleanliness requirements of production or experiments. If specific industry operation details (such as the GMP standards for the pharmaceutical industry) are needed, further targeted guidance can be provided.

German
French
Italian
Portuguese
Japanese
Russian